Clinical experience with the manley multivent: assessment of the effect of the ventilator on cardiopulmonary stability in patients under balanced anaesthesia.
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
University of Dar es Salaam
Abstract
The effects of controlled ventilation by the Manley
Multivent Ventilator on the cardiopulmonary stability in
patients who underwent various surgical procedures under
balanced anaesthesia, were studied in 102 patients of
American Society of Anaesthesiologists (ASA) class 1 and 2.
'The general performance of the ventilator and complications
arising from its use were also assessed.
Controlled ventilation by intermittent positive pressure
ventilation (IPPV) was found to have no significant effect
(P > 0.10) on the blood pressures of the patients. The
pulse rate changes were also not significant (P > 0.05) i
and a good peripheral oxygen saturation of 96 to 100% was
maintained in all patients. The Multivent performed
satisfactorily to provide effective ventilation using
maximum inspiratory pressures of 20 to 35cmH20, tidal
volumes of 6.7 to 12.5mls/kg body wt., respiratory rate 12
or 14 cycles/min and I:E ratio of 1:2. These parameters
were regulated and adjusted successfully according to
individual patient requirements, and adequate and safe
ventilation was achieved. The ventilator was simple to
operate, versatile, reliable and safe for use during
anaesthesia and no complication was associated with its
use.
The Multivent with its unique features of flexibility in
both power and gas supply, coupled with its low purchasing
iii
and running costs, provides anesthetist in the developing
world with an economical, reliable and safe means of
controlled ventilation.